Setting Institutional Defaults (Sample)
N-550
When the desired option is indicated in the %SpO
SatSeconds Alarm Limit button to set the current selection. The N-550 sounds
a confirmation tone.
Set alarm limits to the values established by your facility. After all alarm
limits are set, you will access menu Item 9 to set the alarm limit values to
institutional defaults. The following default values may be set:
Alarm Suspend (30, 60, 90, 120 seconds)
Alarm Silence Behavior (silence off, sound reminder, do not allow
OFF)
Alarm Volume (1 to 10)
Baud Rate (2400, 9600, 19200)
Data Port Mode (ASCII)
Pulse Beep Volume (0 to 10)
Pulse Rate Upper Alarm Limit (lower limit plus one to 250 bpm)
Pulse Rate Lower Alarm Limit (30 bpm to upper limit minus one)
SatSeconds (OFF, 10, 25, 50, 100)
SpO
Upper Alarm Limit (lower limit plus one to 100%)
2
SpO
Lower Limit (20% to upper limit minus one)
2
1. Disconnect the sensor from the N-550.
2. Press the Power On/Off button to turn the N-550 off.
3. Press the Power On/Off button to turn the N-550 on.
4. Set the desired parameters to the institutional values.
Audible Alarm Settings and Service Functions
display, press the
2
41