Setting Institutional Defaults (Sample)
N-550
When the desired option is indicated in the %SpO
SatSeconds Alarm Limit button to set the current selection. The N-550 sounds
a confirmation tone.
Set alarm limits to the values established by your facility. After all alarm
limits are set, you will access menu Item 9 to set the alarm limit values to
institutional defaults. The following default values may be set:
•
Alarm Suspend (30, 60, 90, 120 seconds)
•
Alarm Silence Behavior (silence off, sound reminder, do not allow
OFF)
•
Alarm Volume (1 to 10)
•
Baud Rate (2400, 9600, 19200)
•
Data Port Mode (ASCII)
•
Pulse Beep Volume (0 to 10)
•
Pulse Rate Upper Alarm Limit (lower limit plus one to 250 bpm)
•
Pulse Rate Lower Alarm Limit (30 bpm to upper limit minus one)
•
SatSeconds (OFF, 10, 25, 50, 100)
•
SpO
Upper Alarm Limit (lower limit plus one to 100%)
2
•
SpO
Lower Limit (20% to upper limit minus one)
2
1. Disconnect the sensor from the N-550.
2. Press the Power On/Off button to turn the N-550 off.
3. Press the Power On/Off button to turn the N-550 on.
4. Set the desired parameters to the institutional values.
Audible Alarm Settings and Service Functions
display, press the
2
41