Performance Verification

Test #3: Modulation Level

24
6. The N-550 will display:
75 %SpO2 (pass criteria is 73 to 77 %SpO
60 bpm
alarm
pulse amplitude indicator - low level modulation
1. Press the SRC-MAX % Modulation selection button. The SRC-MAX %
Modulation
LED will light.
2. The N-550 pulse amplitude indicator will fill (10 bars) with each pulse
beat.
3. The N-550 will display:
75 %SpO2 (test pass criteria is 73 to 77 %SpO
60 bpm (test pass criteria is 57 to 63 bpm)
alarm
pulse amplitude indicator - high level modulation
4. Perform Test #1: BPM on page 22. The pulse amplitude indicator should
indicate high level modulation.
5. Perform Test #2: SpO2 on page 23. The pulse amplitude indicator should
indicate high level modulation.
)
2
)
2