3.2.4

Necessary Tools and Components

Observe the following checklist of tools and components that you will need for installing the
Analyzer and GC Controller:
(1)
Chromatographic grade carrier gas: zero grade helium or nitrogen (99.995% pure, with less
than 5 ppm water, and less than 0.5 ppm hydrocarbons).
(2)
High pressure dual-stage regulator for the carrier gas cylinder, high side up to 3000 pounds
per square inch, gauge (psig), low side capable of controlling pressure up to 150 psig.
(3)
Calibration standard gas with correct number of components and concentrations (see Section
2.1.6, this manual).
(4)
Dual-stage regulator for the calibration gas cylinder, low pressure side capable of controlling
pressure up to 30 psig.
(5)
Sample probe (fixture for procuring the stream, or sample gas for chromatographic analysis).
(6)
1/8-inch stainless steel (SS) tubing for connecting calibration standard to analyzer, 1/4-inch
SS tubing for connecting helium carrier to the analyzer, 1/8-inch SS tubing for connecting
stream gas to the analyzer.
(7)
Miscellaneous Swagelok tube fittings, tubing benders and tubing cutter.
(8)
14 American Wire Gauge (AWG) (18 Metric Wire Gauge) or larger electrical wiring and
conduit to provide 115 or 230 volts AC, single phase, 50 to 60 Hertz (Hz), from an
appropriate circuit breaker and power disconnect switch. (See previous guidelines in Section
3.1.2.)
3-16
Observe Precautions and Warnings
1
Plan Site Location
2
Obtain Supplies and Tools
3
Install Analyzer Wiring
4
MODEL 1000