SEP 2005
1.6.4
Basic Analysis Computations
Two basic analysis algorithms are included in the GC Controller. These are:
Area Analysis - Calculates area under component peak
Peak Height Analysis - Measures height of component peak
Concentration Analysis by Using Response Factor
Concentration calculations require a unique response factor for each component in an analysis.
These response factors may be manually entered by an operator or determined automatically by the
system through calibration procedures (with a calibration gas mixture that has known
concentrations).
Response factor calculation: (using the external standard)
Where:
ARF
=
n
HRF
=
n
Area
=
n
Ht
=
n
Cal
=
n
Calculated response factors are stored by the GC Controller for use in the concentration calculations,
and are printed out in the configuration and calibration reports.
1-16
or
Area response factor for component "n" in area per mole percent (%)
Height response factor for component "n"
Area associated with component "n" in calibration gas
Height associated with component "n" in mole % in calibration gas
Amount of component "n" in mole % in calibration gas
MODEL 1000
DESCRIPTION