Leak test all other sample stream lines by connecting gas to each of the sample streams and
repeating steps (9) through (10)(b).
(12)
Finish the test and set up the Analyzer for normal operation, as follows:
(a)
Ensure that all Analyzer Valve switches, upper purged enclosure, are set to the
AUTO position.
(b)
Unplug, or open the "MV" and "SV" vent lines.
(c)
If the calibration gas bottle was used to leak check the sample stream lines, reconnect
the calibration gas bottle to the calibration gas line on the SCS mounting plate, and
reconnect the sample stream lines.
5.5.4.5

Plugged Lines, Columns, or Valves

To ensure that lines, columns, and valves are not plugged, check the gas flow at valve ports. For a
reference, use the flow diagram in the drawing package, and remember these points about flow
diagrams:
-
Port-to-port flow paths are indicated by solid or dashed lines.
-
A dashed line indicates flow direction when the valve is ON, that is, energized.
-
A solid line indicates flow direction when the valve is OFF, that is, de-energized.
-
A combination of solid and dashed lines indicates a constant flow path regardless of
the ON/OFF state of the valve.