Valving; Calibration Gas - Emerson Daniel Danalyzer 1000 Hardware Reference Manual

Gas chromatograph
Table of Contents

Advertisement

SEP 2005
2.1.5

Valving

A block valve should be installed immediately downstream of the sample takeoff point to permit
shutdown of the system for maintenance. Block valves should be either gate valves or cocks of the
proper material and packing, and should be rated for the process line pressure. Tight seating of all
connections is essential.
2.1.6

Calibration Gas

A calibration gas used for Process analysis should be blended to Primary Standards. Primary
Standards are blended using weights that are traceable to the National Bureau of Standards (NBS).
If the calibration standard is a gas, the standard should not have any component that could drop out
at the coldest temperature to which the gas will be subjected. A typical blend for a temperature of
zero degrees Fahrenheit (0°F) is listed in the following table. No dropout will occur in this
calibration gas if it is blended at a pressure below 250 psig.
If a liquid calibration standard is being used, the head pressure must be sufficient to prevent
bubble-out of components during hot weather.
2-4
Table 2-1. Contents of Example Calibration Gas
Gas
Nitrogen
Carbon dioxide
Methane
Ethane
Propane
Isobutane
N-butane
Neopentane
Isopentane
N-pentane
N-hexane
Mole Percent
2.5
0.5
Balance
5.0
1.0
0.3
0.3
0.1
0.1
0.1
0.03
EQUIPMENT DESCRIPTION
MODEL 1000

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount analytical 1000

Table of Contents