Driver Outputs; General Controller Specifications - Emerson Daniel Danalyzer 1000 Hardware Reference Manual

Gas chromatograph
Table of Contents

Advertisement

SEP 2005
2.3.1.4

Driver Outputs

The GC Controller has eight stream switch outputs, 120 mA continuous current, which can be used
to control optional AC or DC solenoid switch boards. This increases stream switch capability from
the standard capability of four gas streams and 1 calibration gas (CAL) system to a maximum
capability of twelve streams.
The optional stream switching assembly (AC or DC solenoid systems) has provisions for eight
stream routes, but adding this option has the net effect of adding only seven more stream
routes to the standard stream capacity of five (thus giving a total stream capacity of twelve).
The reason is, when the optional stream switch assembly is installed, one of the Analyzer's
standard five stream routes becomes dedicated to the optional stream switch assembly.
2.3.1.5

General Controller Specifications

a.
Power requirements (without current outputs): 63.25VA typical for basic instrument
b.
Voltage options:
1. 115 VAC ±15 percent, 50 to 60 Hz @ 0.33 A
2. 230 VAC ± 15 percent, 50 to 60 Hz @ 0.275 A
c.
Temperature
1. Operating range: -18°C to 55°C (0°F to 131°F)
2. Storage range: -40°C to 85°C (-40°F to 185°F)
d.
Humidity: 0 to 95 percent relative humidity, noncondensing
2-14
MODEL 1000
EQUIPMENT DESCRIPTION

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount analytical 1000

Table of Contents