To Add An Ethernet Card - Emerson Daniel Danalyzer 1000 Hardware Reference Manual

Gas chromatograph
Table of Contents

Advertisement

MODEL 1000
F.6

TO ADD AN ETHERNET CARD

(1)
At the GC Controller site, remove the Controller enclosure's front panel.
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Before removing the unit cover from the GC Controller, make certain the power supply
switch is OFF and the AC power cord is disconnected. Observe all safety precautions
when you are working in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.
(a)
For the purged Controller, the front panel is secured by 16 screws. Remove those
screws first.
(b)
Then carefully lower the front panel on its bottom hinges. The front panel is heavy,
so make sure it does not drop and cause damage.
(c)
For the rack mount Controller, the rear of the enclosure is open; it allows access for
most field wiring procedures without removing the enclosure.
(2)
Locate the GC Controller's Terminal Board for Field Wiring (TB). The TB is attached to the
GC Controller's card cage assembly, facing the enclosure's front panel. (In the rack mount
Controller, the TB faces outward toward the rear of the enclosure.)
(3)
Loosen the six screws that secure the TB. Then unplug the TB from its connections at the
back, top of the board.
(4)
Lower the TB down and out of the way, held in place by its ground straps at the bottom of
the board. This exposes the Card Cage Assembly.
(5)
Loosen the four screws that secure the Card Cage Assembly to the chassis. Then remove the
Card Cage Assembly away from its chassis mount so that it is easy to work on.
2350 to 2350A CPU RETROFIT INSTRUCTIONS
SEP 2005
F-17

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount analytical 1000

Table of Contents