Analyzer Specifications - Emerson Daniel Danalyzer 1000 Hardware Reference Manual

Gas chromatograph
Table of Contents

Advertisement

MODEL 1000
2.2.5

Analyzer Specifications

a.
Power Requirements: 120 volts, alternating current (VAC), +10/-15VAC; 50 to 60 Hertz
(Hz); single phase; 10 amperes (A) (maximum during warmup) (additional 10 amps if unit
has a Sample System Oven).
b.
Ambient Temperature Range: -18 degrees Celsius (°C) to +55°C (0°F to +130°F)
c.
Humidity: 0 to 95 percent relative humidity, noncondensing
d.
Vibration: Designed for mounting on process piping or other field structures subject to
normal process vibrations
e.
National Electrical Code (NEC) Area Classification: Suitable for NEC Class 1, Division 1,
Group D
f.
Frame Size (approximately):
-
Height:
-
Width:
-
Depth:
g.
Weight: Approximately 125 pounds (56.8 kilograms [kg]), including mounting hardware.
h.
Sample Requirements:
1.
Fluid Phase - Vapor
2.
Pressure - 15 to 30 psig, regulated to ±10 percent
3.
Flow Rate - 50 cc/min, typical
i.
Analyzer Output Signal: Four different gain channels to provide a 4 to 20 mA signal to the
Controller.
j.
Transient Over Voltages Category:
Installation Category (Over Voltage Category II)
k.
Cleaning requirements are restricted to the 6-port valve (refer to Section 5.5.5.1 Valve
Cleaning, in the manual).
EQUIPMENT DESCRIPTION
78 inches (198 centimeters [cm])
24 inches (61 cm) maximum
24 inches (61 cm)
SEP 2005
2-9