SEP 2005
5.8.5

Performing an Automated Calibration

To perform an automated analog output calibration, from the Analog Outputs dialog box:
Use the Application > Analog Outputs menu path to access this function.
(1)
(2)
The Analog Outputs dialog box appears.
(3)
Select the desired analog output by clicking anywhere in the corresponding row.
(4)
Click on the AutoCal(F8) button or press the F8 key.
You can abort this process at any time by clicking on the Cancel button.
(5)
The Zero Scale Adjustment dialog box appears ("Current device value on engineering
units").
Input the appropriate value and click on the OK button.
5-42