6.7 Front Case Disassembly Procedures

This section describes the items that may be removed on the front case assembly.
A. Front Case Disassembly
A1. To remove the LCD Bracket from the front case
A1-1. Main PCB Disassembly
N5500 Service Manual
Disassembly Guide
a. Disconnect the Encoder wire, the Function
switch board wire and the Power switch
board wire from the Main PCB.
b. Remove 4 corner screws on the LCD
Bracket.
c. Remove the LCD Bracket from the front
case.
a. Disconnect the Inverter wire and the LED
Board wire from the Main PCB.
b. Remove 4 screws on the Main PCB.
c. Remove the Bridge bracket from the Main
PCB.
d. Remove the LCD cable from the Main
PCB.
51