Teledyne 465M User Manual page 63

Ozone monitor
Table of Contents

Advertisement

9.10.2.1. Upload Firmware
1. Press "Write". The screen will display, "Wait for boot loader".
2. When the screen displays, "Power cycle instrument now", power up the
instrument.
Search for boot loader will time out in 20 seconds. If the instrument
is not powered up in 20 seconds, the screen will display, "No boot
loader detected." Go back to step 1 to rewrite the firmware.
The firmware loader software should display a message "Boot loader found,
uploading...". If this message is not displayed, power down the instrument and
receck the COM port configuration and cable (see Section 9.10.1) and retry.
3. When the upload is complete, a message, "Write-flash completed successfully"
should be displayed. If it is an unsuccessful upload, a message, "Write-flash
error" is displayed. If a Write-flash error occurs, ensure that the correct PC COM
port is selected for the instrument (see Section 9.10.2, step 4).
It is not recommended to stop the writing process half way through.
This will cause the firmware to not be fully uploaded.
9.10.2.2. Verify Firmware
The verify function compares the instrument's current firmware to the selected HEX file
(see Section 9.10.2, step 5) and confirms whether or not the files are a match.
1. Press "Verify". The screen will display, "Wait for boot loader".
2. When the screen displays, "Power cycle instrument now", power up the
instrument.
Search for boot loader will time out in 20 seconds. If the instrument
is not powered up in 20 seconds, the screen will display, "No boot
loader detected." Go back to step 1 to verify the firmware.
The firmware loader software should display a message "Boot loader found,
verifying...". If this message is not displayed, power down the instrument and
receck the COM port configuration and cable (see Section 9.10.1) and retry.
06160J DCN7969
NOTE
NOTE
NOTE
Teledyne API Model 465M O
Monitor User Manual
3
61