6. Remove the fours screws, the four washers and the sample filter body from the
sample filter base.
7. Remove the o-ring and inspect for any cracks or deformities. If there are no cracks
or deformities, place the o-ring back in the sample filter body.
If the o-ring is cracked or deformed, discard the o-ring and
replace with a new one.
8. Remove the two sample filter retainers and glass fiber filter element from the
sample filter body.
9. Discard the glass fiber filter element.
10. Place a new glass fiber filter element in between the two sample filter retainers
and place in the sample filter body.
11. Place the sample filter body on the sample body base and secure with the four
screws and fours washers.
12. Reinstall the top cover on the instrument and secure with the six screws or close
the front panel.
54
Teledyne API Model 465M O
NOTE
Monitor User Manual
3
06160J DCN7969