View Menu; Figure 7-4: Span Cal Menu; Table 7-5: View Menu Parameters - Teledyne 465M User Manual

Ozone monitor
Table of Contents

Advertisement

Enter Span target
concentration (actual ozone
concentration being supplied
to monitor)
SPAN CAL
ENT
7.2.4.

VIEW Menu

The VIEW menu allows the operator to view various measurement parameters in real-
time. This can be useful for diagnosing various instrument or system problems.
These values are updated in real-time as they are repeatedly
scrolled on the display.
Parameter
MEAS
UV detector reading, measure cycle
REF
UV detector reading, reference cycle
PRESSURE
Sample pressure
LAMP
UV lamp temperature
TEMP
SAMPLE
Sample temperature
TEMP
FLOW
Sample Gas Flow Rate**
*After initial 20 minute warm-up period.
** Optional Hardware.
06160J DCN7969
ENT TO
900
ENT
SPAN
CFG

Figure 7-4: Span Cal Menu

NOTE

Table 7-5: VIEW Menu Parameters

Meaning
Teledyne API Model 465M O
SUCCESSFUL?
ENT
No
OUT OF
RANGE
ENT
Units
Normal Range*
mV
mV
psia
ºC
ºC
ON/OFF
Monitor User Manual
3
RETURN TO
Yes
CONC MODE
250 – 1230
250 – 1230
9.0 – 16.0
50.0 – 52.0
5.0 –50.0
ON
39