Replacing Internal Particulate Filter; Figure 9-2: Instrument Layout, Nema Configuration - Teledyne 465M User Manual

Ozone monitor
Table of Contents

Advertisement

DRYER

Figure 9-2: Instrument Layout, NEMA Configuration

9.3.

Replacing Internal Particulate Filter

1. Disconnect power from the 465M.
2. Loosen the six screws from the top cover (Rack Mount Configuration, see Figure
9-1) or open front panel (NEMA Configuration, see Figure 9-2.)
3. Remove the six screws and top cover from the instrument.
4. Locate the particulate housing filter on the rear panel (Rack Mount Configuration,
see Figure 9-1) or the bottom panel (NEMA Configuration, see Figure 9-2.)
5. Loosen the four screws on the sample filter body (see Figure 9-3.)
06160J DCN7969
Teledyne API Model 465M O
Monitor User Manual