Figure 4-2. Nema Ac Power Connection To Monitor - Teledyne 465M User Manual

Ozone monitor
Table of Contents

Advertisement

3. For the NEMA configuration, the electrical connection must be hard-wired to the 3-
pin connector on the mainboard labeled J23. AC power connection to the monitor
should be made with 12-14 AWG stranded copper wire, connected to the monitor as
follows:
a. Earth Ground (green): Connect the earth ground wire to the screw terminal lug on
b. Line (Black):
c. Neutral (White):

Figure 4-2. NEMA AC Power Connection to Monitor

4. Connect the power cord to an appropriate power outlet (see the serial number tag for
correct voltage and frequency).
Verify that the instrument is set up for proper line voltage and
frequency. Observe local electrical codes when connecting
18
Teledyne API Model 465M O
the chassis as shown in the Figure below. Do not connect
the earth ground wire directly to the PCB connector on the
mainboard PCB.
Connect the Line wire to the connector on the mainboard
PCB labeled "Line".
Connect the Neutral wire to the connector on the
mainboard PCB labeled "Neutral".
WARNING: Electrical Shock Hazard
power to the monitor.
Monitor User Manual
3
06160J DCN7969