Outputs; Table 10-1: Status Led/Output Definitions - Teledyne 465M User Manual

Ozone monitor
Table of Contents

Advertisement

10.2.
Troubleshooting Using Front Panel Status
LED's or Status Outputs
The 465M has 4 front-panel status LED's that reflect the current operating status of the
monitor, and indicate fault conditions. There are also four relay Status Outputs on the
rear panel (or internally in the NEMA configuration) that also reflect the state of these
status LED's. See Section 4.5.3 for more information on the Status Outputs.

Table 10-1: Status LED/Output Definitions

Status
Status LED
Output #
Label
1
Sensor OK
2
Invalid
Reading
3
Check Lamp
4
Pneumatic
Error
*If Flow Switch installed
06160J DCN7969
Normal
Triggers
Operating
State
• Reference < 125.0 mV
On
• Reference > 1230.0 mV – 0.5 Wt%
• O3 Concentration < F.S. Range
Off
• O3 Concentration > F.S. Range
• Reference < 250.0 mV
Off
• Pressure < 9.0 psia
Off
• Pressure > 16.0 psia
• Flow < 500 cc/min*
Teledyne API Model 465M O
Monitor User Manual
3
Critical
Warning?
Yes
No
No
No
65