Service Manual

Clean the Air Filter

To prevent risk of injury, do not operate or power the Product
without the fan filter in place.
Damage can occur when the Product becomes too hot if the
area around the fan is restricted, the intake air is too warm, or
the filter becomes clogged.
The air filter must be removed and cleaned at 30 day intervals or more frequently
if the Product is operated in a dusty environment. Access the air filter from the
rear panel of the Product.
To clean the air filter, refer to Figure 25 and continue as follows:
1. Turn the power off and let the fan come to a stop.
2. Disconnect the ac line cord.
3. Remove the filter element.
4. Clean the filter element,
5. Do the filter removal steps in reverse order to install the filter element.
84
a. Hold the top and bottom of the air filter frame.
b. Squeeze the edges of the frame towards each other to disengage the
filter tabs from the slots in the Product.
c. Pull the filter frame straight out from the Product.
a. Clean the filter element in soapy water.
b. Flush the filter element.
c. Shake out the water and let the filter element dry before you install it.
W Warning
W Caution