CAL_DAYS?
Description Gets the number of days and hours since the last calibration
Parameters MAIN
Example
Response (Integer) Days.
CAL_FACT
Description Set the procedure "fault action" flag. Procedures refer to calibration
Parameters CONT
Example
CAL_FACT?
Description Get the procedure "fault action" flag.
Example
Response ABORT.
CAL_FAULT?
Description Get information about calibration error, if one occurred).
Example
Response 1. Error number (use EXPLAIN? command to interpret).
CAL_INFO?
Description Return message or instructions related to the running step.
Example
Response (String) the message string.
CAL_NEXT
Description Continue a calibration procedure if it is waiting for a CAL_NEXT
Parameters (Optional) reference value (used if it's waiting for a reference). If the
Example
76
constant. The data output for date is in the same format as the
CLOCK command.
Returns the days since the main unit calibration.
ZERO
Returns the days since the ZERO calibration.
OHMSZERO
Returns the days since the Ohms zero calibration.
SCOPE
Returns the days since the scope option calibration.
CAL_DAYS? ZERO
(Integer) Hours.
and diagnostic procedures. This command is more useful for
diagnostics that calibration.
To continue on faults.
ABORT
To abort on faults.
CAL_FACT ABORT
CAL_FACT?
CAL_FAULT?
2. Name of step where error occured.
CAL_INFO?
command.
reference value has no unit, the unit is assumed to be that returned
by the CAL_REF? command.
CAL_NEXT