Fluke 5080A Service Manual page 88

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

5080A
Service Manual
CAL_STORE?
Description Gets if a CAL_STORE is necessary.
Response 1
Example
CAL_SW?
Description Gets the position of the calibration switch.
Response 1
Example
EOFSTR
Description Sets the End-Of-File character string used for calibration reports.
Parameters The EOF string (two characters maximum).
EOFSTR?
Description Gets the End-Of-File character string used for calibration reports.
Parameters None
Response (String) The End-Of-File character string.
PR_RPT
Description Prints a self-calibration report out the selected serial port.
Parameters 1. Type of report to print: STORED, ACTIVE, or CONSTS
Example
RPT?
Description Gets a self-calibration report.
Parameters 1. Type of report to print: STORED, ACTIVE, or CONSTS
Example
RPT_PLEN
Description Sets the page length used for calibration reports. This parameter is
Parameters Page length.
78
Yes.
No
0
CAL_STORE?
Enabled.
Normal
0
CAL_SW?
The maximum length is two characters. The EOF parameter is kept
in nonvolatile memory.
2. Format of report: PRINT (designed to be read) SPREAD
(designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report: I90D (90 day specifications) or I1Y (1 year specifications)
4. Serial port out which to print report: HOST or UUT
PR_RPT STORED,PRINT,I90D,HOST
2. Format of report: PRINT (designed to be read) SPREAD
(designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report: I90D (90 day specifications) or I1Y (1 year specifications)
PR_RPT? STORED,PRINT,I90D
kept in nonvolatile memory.