Fluke 5080A Service Manual page 62

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

5080A
Service Manual
52
Table 37. Calibration Steps for AC Current
Steps
Amplitude
1
3.2999 mA
2
0.3300 mA
3
3.0000 mA
4
3.00000 mA
5
300.0uA
6
300.0uA
7
30.000 mA
8
30.000 mA
9
300.000 mA
12
300.000 mA
13
1.9000 A
15
1.9000 A
16
10.0000 A
17
10.0000 A
Output (AUX HI, LO)
Frequency
55.00 Hz
Fluke A40 10 mA
55.00 Hz
Fluke A40 10 mA
55.00Hz
Fluke A40 10 mA
1.000 kHz
Fluke A40 10 mA
55.00 Hz
Fluke A40 10 mA
1.000 kHz
Fluke A40 10 mA
55.00 Hz
Fluke A40 200 mA
1.000 kHz
Fluke A40 200 mA
55.00 Hz
Fluke A40 2 A
1.000 kHz
Fluke A40 2 A
55.00 Hz
Fluke A40 2 A
1.000 kHz
Fluke A40 2 A
AUX 20A, LO
55.00 Hz
Fluke A40A 20 A
1.000 kHz
Fluke A40A 20 A
Shunt to Use

Advertisement

Table of Contents
loading

Table of Contents