Table 42. Required Test Equipment for SC200 Calibration (cont.)
Recommended Model
Edge Rise Time and Aberrations Verification
Tektronix 11801 with SC-
22/26 sampling head, or
Tektronix TDX 820 with
8 GHz bandwidth
Weinschel 9-10 (SMA) or
Weinschel 18W-10 or
equivalent
(supplied with SC200)
Leveled Sine Wave Amplitude Calibration and Verification
Fluke 5790A
Pomona #1269
(supplied with SC200)
DC and AC Voltage Calibration and Verification, DC Voltage Verification
Hewlett Packard 3458A
Pomona #1269
(supplied with SC200)
Leveled Sine Wave Frequency Verification
PM 6680 with Option (PM
9621, PM 9624, or
PM 9625) and (PM 9678)
Pomona #3288
(supplied with SC200)
Leveled Sine Wave Flatness (Low Frequency) Calibration and Verification
Fluke 5790A with 03
option
Pomona #3288
(supplied with SC200)
Leveled Sine Wave Harmonics Verification
Hewlett Packard 8590A
Pomona #3288
(supplied with SC200)
Minimum Specifications
Frequency
2 GHz
Resolution
4.5 mV to 2.75 V
10 dB, 3.5 mm(m/f)
BNC(f) to 3.5 mm(m)
Range
5 mV p-p to 5.5 V p-p
Frequency
50 kHz
BNC(f) to Double Banana Plug
Feedthrough 50 Ω, +1 %
BNC(f) to Double Banana Plug
Feedthrough 50 Ω, +1 %
50 kHz to 350 MHz, <1.6 ppm uncertainty
BNC(f) to Type N(m)
Range
5 mV p-p to 5.5 V p-p
Frequency
50 kHz to 10 MHz
BNC(f) to Type N(m)
BNC(f) to Type N(m)