Calibrate Through the Remote Interface

To access the standard calibration steps through the remote interface, send the
command:
To jump to a specified calibration step, add a comma followed by a modifier to
specify an entry point. Table 50 shows the applicable entry points and the
modifier to use for each point.
For example, to jump directly to ac volts calibration, send the command:
To go directly to Resistance calibration, send the command:
These calibration commands can be sent through the IEEE-488 or serial
interface. To type in commands through the serial interface:
1. Use a Fluke PM8914 cable to connect the applicable COM port on a PC to
2. In Microsoft Windows, start the Terminal program. Set the communications
3. Push . At the prompt, type the calibration command, for example
74
CAL_START MAIN
Table 50. Jumping to a Specific Calibration Step in Remote
Entry points for CAL_START MAIN
AC Volts
Thermocouple Measuring
DC Current
AC Current
AUX DC Volts
AUX AC Volts
Resistance
Capacitance
Entry Points for CAL_START FACTORY
NORMAL Volts and AUX Volts Phase
Volts and Current Phase
CAL_START MAIN,AV
CAL_START MAIN,R
the 5080A Serial 1 connector.
parameters to agree with that of the Product.
CAL_START MAIN.
Modifier
AV
TEMPX
ICAL
AI
V2
AVS
R
C
Modifier
PHASE
IPHASE