5080A
Service Manual
Low Resistance Source (LVR) Verification
To verify the Low Resistance Source function:
1. Connect the low resistance source output to a standard multimeter as shown
2. Push on the Product.
3. Push the softkey labeled MODE until lvr shows above the right-most Product
4. Set the multimeter for true ohms and auto range.
5. Measure the Low Resistance calibration performance at each resistance
1.0000 Ω
1.8000 Ω
3.7000 Ω
5.9000 Ω
10.000 Ω
18.000 Ω
37.000 Ω
59.000 Ω
100.00 Ω
180.00 Ω
370.00 Ω
590.00 Ω
1.0000 kΩ
1.8000 kΩ
3.7000 kΩ
5.9000 kΩ
30
in Figure 7. Use a 4-wire connection and setup the multimeter for a 4-wire
ohms measurement.
Fluke 5080A
Figure 7. Low Resistance Verifications Connections
softkey.
point in Table 15. Deviations must not be larger than the specified limits.
Table 15. Megohm Option LVR Verification Points
Range
Output
1.0000 Ω
1.8000 Ω
3.7000 Ω
5.9000 Ω
10.000 Ω
18.000 Ω
37.000 Ω
59.000 Ω
100.00 Ω
180.00 Ω
370.00 Ω
590.00 Ω
1.0000 kΩ
1.8000 kΩ
3.7000 kΩ
5.9000 kΩ
Fluke 8508A
gru001.eps
Tolerance
±0.0109 Ω
±0.0139 Ω
±0.0208 Ω
±0.0288 Ω
±0.045 Ω
±0.075 Ω
±0.154 Ω
±0.281 Ω
±0.45 Ω
±0.75 Ω
±1.5 Ω
±2.0 Ω
±0.003 kΩ
±0.004 kΩ
±0.005 kΩ
±0.006 kΩ