Ac Volts Calibration (Normal Output) - Fluke 5080A Service Manual

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

5080A
Service Manual
Fluke 8508A
Set the Fluke 8508A
to external guard

AC Volts Calibration (NORMAL Output)

The equipment shown in Table 32 is necessary for calibration of the ac volts
function. (The equipment is also listed in the consolidated table, Table 29.)
To calibrate the ac voltage function:
1. Measure the Product output through Input 1 of a Fluke 5790A or 5790B AC
2. Type in the measured values into the Product for each step in Table 33 as
AC Voltage adjust completed.
46
Figure 10. DC Voltage Adjustment Setup
Table 32. Test Equipment Required for AC Volts Calibration
Quantity
Manufacturer
1
Fluke
1
Fluke
Measurement Standard. Use a Type N to dual banana adapter as shown in
Figure 11..
prompted.
Model
PN 900394
Type N to dual banana adapter
5790A, 5790B
AC Measurement Standard
UUT
gru005.eps
Equipment