Detailed Specifications; Dc Voltage - Fluke 5080A Service Manual

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

5080A
Service Manual
International ........................................................ IEC 61326-1: Basic Electromagnetic Environment
Korea (KCC) ....................................................... Class A Equipment (Industrial Broadcasting & Communication
USA (FCC) .......................................................... 47 CFR 15 subpart B. This product is considered an exempt device per
Line Power
Line Voltage (selectable) .................................... 100 V, 120 V, 220 V, 240 V
Line Frequency ................................................... 47 to 63 Hz
Line Voltage Variation ......................................... ±10 % about line voltage setting
Power Consumption................................................ 600 VA
Dimensions
Height.................................................................. 19.3 cm (7.6 in)
Width ................................................................... 43.2 cm (17 in), 44.3 cm (17.5 in) including handles
Depth .................................................................. 53.8 cm (21.2 in)
Weight ..................................................................... 22 kg (48 lb)
Specification Definition ............................................ The specifications include stability, temperature coefficient, linearity,
Specification Confidence Level ............................... 99 %

Detailed Specifications

DC Voltage

Range
0 to 329.999 mV
0 to 3.29999 V
0 to 32.9999 V
10 to 101.999 V
30 to 329.999 V
100 to 1020.00 V
0 to 329.99 mV
0.33 to 3.2999 V
3.3 to 7.000 V
Remote sensing is not provided. Output resistance is 60 Ω for outputs <330 mV. Output resistance is <5 mΩ for outputs ≥0.33 V.
[1]
The AUX output has an output resistance of <1 Ω.
6
CISPR 11: Group 1, Class A
Equipment)
clause 15.103.
line and load regulation, and the traceability of the external standards
used for calibration. It is not necessary to add anything to determine
the total specification for the temperature range indicated.
Specification, tcal ±5 °C
±(% of output + μV)
90 days
1 year
0.011 % + 10
0.013 % + 10
0.008 % + 15
0.010 % + 15
0.008 % + 150
0.010 % + 150
0.010 % + 1500
0.012 % + 1500
0.010 % + 1500
0.012 % + 1500
0.010 % + 5500
0.012 % + 5500
Auxiliary Output (dual output mode only)
0.10 % + 1000
0.12 % + 1000
0.10 % + 1000
0.12 % + 1000
0.10 % + 1000
0.12 % + 1000
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for
the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low-voltage power
supply network that supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated
disturbances.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user should take
notice of it. This equipment is intended for use in business
environments and not to be used in homes.
Stability
24 hours, ±1 °C
±(% of output + μV)
0.0035 % + 6
0.0025 % + 10
0.0025 % + 100
0.003 % + 1000
0.003 % + 1000
0.003 % + 5000
0.03 % + 300
0.03 % + 300
0.03 % + 300
Resolution (μV)
Max Burden
1
60 Ω
10
300 mA
100
600 mA
1000
300 mA
1000
120 mA
10000
40 mA
10
5 mA
100
5 mA
1000
5 mA
[1]

Advertisement

Table of Contents
loading