Table 42. Required Test Equipment for SC200 Calibration (cont.)
Test Equipment
Recommended Model
Edge Frequency, AC Voltage Frequency Verification
PM 6680 with PM 9678
Frequency Counter
option
BNC Cable
(supplied with SC200)
Frequency Counter
PM 6680
BNC Cable
(supplied with SC200)
Leveled Sine Wave Flatness (High Frequency) Calibration and Verification
Power Meter
Hewlett Packard 437B
Power Sensor
Hewlett Packard 8482A
Power Sensor
Hewlett Packard 8481A
30 dB Reference
Hewlett Packard 11708A
Attenuator
(supplied with HP 8481D)
Hewlett Packard
Adapter
PN 1250-1474
BNC Cable
(supplied with SC200)
Leveled Sine Wave Frequency, Time Marker Verification
PM 6680 with Option (PM
Frequency Counter
9621, PM 9624, or
PM 9625) and (PM 9678)
Adapter
Pomona #3288
BNC Cable
(supplied with SC200)
AC Measurement
Fluke 5790A or 5790B
Standard
with 03 option
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC200)
20 ms to 150 ns, 10 Hz to 10 MHz:
<1.6 ppm uncertainty
Edge Duty Cycle
Range
Frequency
Range
Frequency
Range
Frequency
Range
Frequency
BNC(f) to Type N(f)
2 ns to 5 s, 50 kHz to 500 MHz: <1.6 ppm
uncertainty
BNC(f) to Type N(m)
Wave Generator Verification
Range
Frequency
BNC(f) to Double Banana Plug
Feedthrough 50 Ω, +1 %
Minimum Specifications
-42 dBm to +5.6 dBm
10 MHz – 300 MHz
-20 dBm to +19 dBm
10 MHz – 300 MHz
-42 dBm to -20 dBm
10 MHz – 300 MHz
30 dB
50 MHz
1.8 mV p-p to 55 V p-p
10 Hz to 100 kHz
Calibrator
SC200 Calibration
59