Table 8. Verification Tests for AC Voltage (Normal) (cont.)
Range
Output
3.3 V
3.2900 V
33 V
32.900 V
33 V
32.900 V
33 V
32.900 V
330 V
100.00 V
330 V
100.00 V
330 V
100.00 V
1000 V
1020.0 V
1000 V
1020.0 V
1000 V
1020.0 V
Table 9. Verification Tests for AC Voltage (AUX)
Range
Output, AUX
330 mV
10.00 mV
330 mV
10.00 mV
330 mV
10.00 mV
330 mV
329.00 mV
330 mV
329.00 mV
330 mV
329.00 mV
3.3 V
3.29000 V
3.3 V
3.29000 V
3.3 V
3.29000 V
5 V
5.00000 V
5 V
5.00000 V
5 V
5.00000 V
Table 10. Verification Tests for AC Current
Range
Output
330 μA
29.0 μA
330 μA
29.0 μA
330 μA
29.0 μA
330 μA
329.0 μA
330 μA
329.0 μA
330 μA
329.0 μA
Frequency
Lower Limit
1 kHz
3.28653 V
45 Hz
100 Hz
32.8620 V
1 kHz
32.8620 V
45 Hz
100 Hz
1 kHz
45 Hz
1018.60 V
100 Hz
1018.49 V
1 kHz
1018.49 V
Frequency
Lower Limit
45 Hz
100 Hz
1 kHz
45 Hz
327.408 mV
100 Hz
327.342 mV
1 kHz
327.342 mV
45 Hz
3.283078 V
100 Hz
3.282420 V
1 kHz
3.282420 V
45 Hz
4.990000 V
100 Hz
4.989000 V
1 kHz
4.989000 V
Frequency
Lower Limit
45 Hz
100 Hz
1 kHz
327.46 μA
45 Hz
327.43 μA
100 Hz
327.43 μA
1 kHz
Upper Limit
3.29347 V
32.8686V
32.9314V
32.9380 V
32.9380 V
99.862 V
100.138 V
99.852 V
100.148 V
99.862V
100.138V
1021.40 V
1021.51 V
1021.51 V
Upper Limit
8.982 mV
11.018 mV
8.980 mV
11.020 mV
8.980 mV
11.020 mV
330.592 mV
330.658 mV
330.658 mV
3.296922 V
3.297580 V
3.297580 V
5.010000 V
5.011000 V
5.011000 V
Upper Limit
28.18 μA
29.82 μA
28.18 μA
29.82 μA
28.18 μA
29.82 μA
330.54 μA
330.57 μA
330.57 μA
Calibrator
Performance Tests
27