Fluke 5080A Service Manual page 63

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

Fluke Calibration Corporation - Worldwide Support Center MET/CAL Procedure
=============================================================================
INSTRUMENT:
DATE:
AUTHOR:
REVISION:
ADJUSTMENT THRESHOLD:
NUMBER OF TESTS:
NUMBER OF LINES:
CONFIGURATION:
=============================================================================
STEP
FSC
RANGE NOMINAL
# 10 Sep 98 changed Cal_Info? commands to Out? and checked for 10A -
# needs cal_next to get past display; check for 0 out when ACI is done.
#
1.001
ASK-
R
Q N
1.002
HEAD
# Set M[10] to 3mA initially
1.003
MATH
# Reset UUT - get it out of calibration mode.
1.004
IEEE
1.005
IEEE
1.006
MATH
1.007
JMPT
1.008
IEEE
1.009
MEME
1.010
JMPZ
1.011
JMP
1.012
HEAD
1.013
DISP
1.013
DISP
1.013
DISP
1.013
DISP
1.013
DISP
1.013
DISP
1.013
DISP
1.014
JMP
# Reset 5790A standard.
1.015
ACMS
1.016
5790
1.017
HEAD
1.018
PIC
1.019
IEEE
1.020
IEEE
1.021
ACMS
1.022
5790
1.023
MATH
# Apply nominal -DC Current to A40
1.024
IEEE
1.025
IEEE
1.026
ACMS
1.027
5790
1.028
MATH
1.029
IEEE
1.030
IEEE
1.031
ACMS
1.032
5790
1.033
MATH
Sub Fluke 5080A ACI ADJ
03-Oct-2011
0.6
70%
1
487
Fluke 5790A
TOLERANCE
U
AC CURRENT ADJUSTMENT
M[10] = 0.003
*CLS;*RST; *OPC?[I]
ERR?[I$][GTL]
MEM1 = FLD(MEM2,1,",")
CAL_SW?[I][GTL]
1.012
1.015
WARNING! CALIBRATION SWITCH IS NOT ENABLED.
The UUT CALIBRATION
switch is in NORMAL.
The switch MUST be in ENABLE to store the
new calibration constants.
Select ENABLE, then press "Advance" to
continue with the calibration process.
1.008
*
*
DCI References
552A410m
OUT 3.2999mA, 0HZ; OPER; *OPC?[I][GTL]
[D30000][GTL]
A
M[17] = MEM
OUT -3.2999mA, 0HZ; OPER; *OPC?[I][GTL]
[D5000][GTL]
A
M[17] = (ABS(MEM) + M[17]) / 2
OUT .33mA, 0HZ; OPER; *OPC?[I][GTL]
[D15000][GTL]
A
M[18] = MEM
Figure 14. Sample MET/CAL Program
MOD1
MOD2
3
4 CON
C
F
S
G
SH
N
G
SH
N
G
SH
N
Calibrator
Mainframe Calibration
W
2W
2W
2W
53