2. Details of Menu Items in Adjustment Settings for Skilled Operators
0505: [Developer Fill: Result]
Display the result of filling developer.
If the developer replenishment fails, check the displayed code and troubleshoot the problem
accordingly. If countermeasures do not work or a code that is not on the list is displayed, contact your
service representative.
For details about filling/discharging developer, see the Replacement Guide.
Code
1
Succeeded (developer
replenished)
2
Developer not discharged
3
Developer not replenished
9
Replenishment quit forcefully
0506: [Developer Exhaust]
Discharge used developer into the developer bottle.
Use this to replace the developer. This operation can take up to 3 minutes. Do not pull the unit out during
the operation. For details about discharging developer, see the Replacement Guide.
[Execute Developer Exhaust]
48
Status
-
Developer in the developer
unit has not been
discharged.
Setting Items
Causes
• The developer bottle is
not attached to the
inlet.
• The developer bottle is
attached to the inlet,
but its seal has not
been pulled out.
• The front covers have
been closed.
• The power was turned
off during the
operation.
Press [OK] to apply the setting(s).
Solutions
-
A new developer bottle is
required. Contact your
service representative.
Attach a new developer
bottle to the inlet, pull out
the seal, and then try the
operation again.
Open the front covers, turn
on the main power, and
then try the operation again.