Table 7-1 Events Which Result In A Change Of Status - Siemens SITRANS P300 Operating Instructions Manual

With hart communication
Table of Contents

Advertisement

Operating functions via HART
7.4 Selecting the measurement type
Table 7-1
Events which result in a change of status
DV
Measurement
BAD
type
0
Pressure
DV3=BAD, fault
in linearization
1
Sensor
DV2 = BAD,
temperature
analog-to-digital
converter in
over/underrange
, fault in
linearization
2
Electronics
Analog-to-digital
temperature
converter in
over/underrange
, fault in
linearization
3
Pressure
Analog-to-digital
(untrimmed)
converter in
over/underrange
, sensor
breakage DV1,
DV2=BAD, fault
in linearization
4
Level
If DV0 = BAD
5
Volume
DV0 = BAD
Characteristic
curve is faulty
6
Mass
DV5 = BAD
7
Volume of
DV0 = BAD
flow
Characteristic
(not absolute
curve is faulty
and gauge
pressure)
8
Mass flow
DV5 = BAD
rate
(not absolute
and gauge
pressure)
7-12
MANUAL
UNCERTAIN
When DV0
DV3 =
is simulated
UNCERTAIN
When DV1
DV1 more than
is simulated
20% outside the
sensor limits
DV2=
UNCERTAIN
DV2 = MANUAL
When DV2
DV2 more than
is simulated
20% outside the
sensor limits
-
Analog-to-digital
converter in
overrange/under
range, DV3
more than 20%
outside the
sensor limits
DV2 = MANUAL
When DV0
DV0=
is simulated
UNCERTAIN
When DV0
DV0=
is simulated
UNCERTAIN
Input value is
outside the
specified
characteristic
curve range
When DV0
DV5=
is simulated
UNCERTAIN
When DV0
DV0=
is simulated
UNCERTAIN
Input value is
outside the
specified
characteristic
curve range
When DV0
DV5=
is simulated
UNCERTAIN
CONSTANT
HIGH LIMIT
-
DV3 = HIGH
LIMIT
-
Analog-to-
digital
converter in
overrange
-
Analog-to-
digital
converter in
overrange
-
Analog-to-
digital
converter in
overrange
DV not active
DV0 = HIGH
LIMIT
Characteristic
DV4 = HIGH
curve is faulty
LIMIT
DV not active
Characteristi
c curve for
maximum
value with
gradient 0
DV not active,
DV5 = HIGH
LIMIT
DV5=CONST
ANT
Characteristic
DV4 = HIGH
curve is faulty,
LIMIT
DV not active
Characteristi
c curve for
maximum
value with
gradient 0
DV not active
DV5 = HIGH
LIMIT
DV5 =
CONSTANT
SITRANS P300 with HART communication
Operating Instructions, 06/2005, A5E00359579-02
LOW LIMIT
DV3 = LOW
LIMIT
Analog-to-
digital converter
in underrange
Analog-to-
digital converter
in underrange
Analog-to-
digital converter
in underrange
DV0 = LOW
LIMIT
DV4 = LOW
LIMIT
Characteristic
curve for
minimum value
with gradient 0
DV5 = LOW
LIMIT
DV4 = LOW
LIMIT
Characteristic
curve for
minimum value
with gradient 0
DV5 = LOW
LIMIT

Hide quick links: