Siemens SITRANS P300 Operating Instructions Manual page 123

With hart communication
Table of Contents

Advertisement

Operating conditions
Installation conditions
Ambient temperature
Observe the temperature class in areas subject
to explosion hazard.
Climatic class
Degree of protection
As per EN 60 529
Electromagnetic compatibility
Process conditions
Medium temperature
Design
Weight
Material
Wetted parts materials
Non-wetted parts materials
Measuring cell filling
Process connection
Electrical connection
SITRANS P300 with HART communication
Operating Instructions, 06/2005, A5E00359579-02
9.2 SITRANS P300 gauge pressure / absolute pressure (flush-mounted)
Measuring cell with silicone oil
Digital display
Storage temperature
Condensation
Spurious emission and resistance to
interference
Measuring cell with silicone oil
Process connection
Seal diaphragm
Electronics housing
-40 °C ... +85 °C (-40 °F ... 185 °F)
-30 °C ... +85 °C (-22 °F ... 185 °F)
-50 °C ... +85 °C (-58 °F ... 185 °F)
permitted
IP65, IP68, NEMA 4X, enclosure cleaning,
resistant to lyes, steam to 150 °C
As per EN 61 326 and NAMUR NE 21
-25 °C ... +150 °C (-13 °F ... 302 °F)
-25 °C ... +200 °C (-13 °F ... 392 °F) with
temperature decoupler
Approx. 1 ... 13 kg (2.2 ... 29 lb)
Stainless steel, mat. no. 1.4404/316L
Stainless steel, mat. no. 1.4404/316L
Electropolished stainless steel, mat. no. 1.4301
Silicone oil
Neobee M20
Medical white oil
Flanges as per EN and ASME
F&B and pharmaceutical flanges
Bioconnect/Biocontrol
PMC Style
Cable entry via threaded connection M20x1.5
(plastic)
Cable entry via threaded connection M20x1.5
(metal with screening)
Specifications
9-7

Hide quick links: