33.8 Electromagnetic Immunity

The System is intended for use in the electromagnetic environment specified below. Always make sure that the System is used in
such an environment.
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Immunity Test
Electrostatic Discharge (ESD) IEC
61000-4-2
Electrical Fast Transient/burst IEC
61000-4-4
Surge IEC 61000-4-5
IEC 60601 Test Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/ output lines (100
kHz repetition frequency)
± 1 kV differential mode
± 2 kV common mode
CHAPTER 33 • Technical Specifications
Electromagnetic Environment –
Compliance Level
Guidance
Floors should be wood, concrete or
± 8 kV contact
ceramic tile. If floors are covered with
± 15 kV air
synthetic material, the relative humidity
should be at least 30%.
± 2 kV for power supply lines
Mains power quality should be that of a
± 1 kV for input/output lines
typical commercial or hospital
(100 kHz repetition frequency)
environment.
Mains power quality should be that of a
± 1 kV differential mode
typical commercial or hospital
± 2 kV common mode
environment.