The FACTOR can be programmed within the
range -2.0000 to +2.0000. For factors below 0.2
or above 2.0 an appropriate decimal point set-
ting should be used by entering the EXPO-
NENT. Using EXPONENT the decimal point
can be shifted as far to the left (-) or to the right
(+) as the display and printer permit. An expo-
nential view of measured values is not possible.
To calculate the scaling values automatically :
5 Base value:
5 Factor, Exponent :
the function menus (see 11.4) include, derived
from the actual values and setpoints, a special
menu
.
SCALING
Once the scaling values have been programmed and the actual measured val-
ue thus modified the correction arrow appears indicating the measured value
status (see 8.2).

12.7 Correction values

Sensors can be corrected by means of the correction values ZERO-POINT
and GAIN; (see Manual 6.3.10).
Corrected measured value = (measured value - ZERO-POINT ) x GAIN
Function
Zero-point correction:
Gain correction:
To switch on and off press :
Once the scaling values have been programmed and the actual measured val-
ue thus modified the correction arrow appears indicating the measured value
status (see 8.2).
To reach maximum accuracy multi-point calibration of sensors is now
also possible - with option KL (see 12.11).

12.8 Changing the units

For each measuring channel the default units for the measuring range can be
replaced with any two-character units; (see Manual 6.3.5). All upper-case and
lower-case letters, special characters °, Ω, %, !, [, ], *, -, =, ~ and space (_) can
be used. The units are shown as two characters after the measured value or
programming value.
To change the units use the function:
------
------E0
Scaling, Decimal point setting
*SENSOR PROGRAMMING 2 *
Connector : 0
5 Base value :
5 Factor, Exp :
4 Zero-point :
4 Gain :
1 Range, Units :
Mªª
M
SCALING :
01:
Actual val 1:04.000 2: 20.000
Decimal pt: 1
Setpoint 1:
-100.0 2: 400.0
4 Gain :
5 Base value :
5 Factor :
Mªª
M
4 Zero-point :
4 Gain :
-----°C
or
<OFF>
<ON>
2 Units :
ALMEMO
®
2590
Channel :
00
------ °C
------E0
------ °C
------
NiCr °C
©P
4.67 mA
Units :
°C
-------
720.0 °C
0.3125 E2
FCT
-----°C
°C
49