11. Function menus

11.5.8.3 Output format

The output format (see Manual 6.6.1) defines the print layout for measuring
point scans and for output of the memory. This output format can be pro-
grammed in the function
which all measured values are listed one below the other; there is also the
´Columns´ format listing them next to one another; this provides a clear, easy-
to-understand, and space-saving printout. For this latter format the printer is
switched over automatically to compressed character mode. There is also the
Table format which is suitable for further processing using a spreadsheet pro-
gram; (see print layouts, Manual 6.1).
Output format ´ ´ List meas values one below the other : Output form : List
Output format ´n´ Columns next to one another :
Output format ´t´ Table, semi-colon separated :
In the data logger menu behind the cycle for memory activation ´S´, if not ´U´
The format is indicated by an abbreviation ´n´ or ´t´:

11.5.8.4 Measuring rate, continuous measuring point scan

As and when necessary the measuring rate (conversion rate) for measuring
point scans can be raised in the
(measuring operations per second) (see Manual 6.5).
Semi-continuous measuring point scan
The measuring rate for measuring points is by default semi-continuous; i.e. all
measuring points are continuously scanned but the selected measuring point is
assigned preferred status and is rescanned every 2nd time. This ensures a
constant scanning rate (= half the measuring rate) irrespective of the number of
measuring channels; this is an advantage for analog output or measured value
smoothing but when averaging (M(n) it may lead to incorrect results.
0 M 1 M 2 M 3 M 4 M 5 M 0 M 1 M 2 M 3 M
Continuous measuring point scan
If continuous measuring point scanning is set, all active measuring chan-
nels are scanned equally often at the chosen measuring rate and uninterrupt-
edly one after the other; (see Manual 6.5.1.3). This almost doubles the mea-
suring rate / channel. In both modes all measured values can be saved and
output at any time. With both the following functions continuous saving to
memory and continuous output of measured values can be activated at the
measuring rate.
Function Meas. rate, to modify press key :
Semi-continuous measuring point scan (standard):
Continuous measuring point scan
Continuous saving to memory OFF
Continuous saving to memory, activate
Continuous output OFF
Continuous output, activate :
44
ALMEMO
Output form
Measuring rate
®
2590
. There is the default format ´List´ in
Output form: Columns
Output form: Table
Cycle timer: 00:15:00 Sn
function from 2.5 to 10 mops
<SET>
Meas. rate: 10 mops
<OFF>
<ON>
Saving to memory:
<ON>
Output:
<ON>
Cont:
-
Cont:
Ø
-
Ø
-
Ø