Weight Calibration
One to ten load weights need to be applied to calibrate the scale. Using multiple point
calibration allows the unit to weigh more accurately; by removing undesirable
characteristics of load cells. A typical weight calibration is a three point calibration. This
means three different and optimal loads are applied and entered (not including the zero
point). If you do not conveniently have the three different weights available, you may
also use one or two point calibration. To calibrate with one point, simply turn off the
scale after completing step 13 as listed below. To calibrate with two points, turn off the
scale after completing step 15 as listed below. The CW250 has the capability to apply
and load up to 10 calibration points.
10. The display will show "SAvE" for about 1 second and advance to show "CAP". Press
the PRINT/ ACCUMULATE button. The display will show the current capacity
setting with the far right number flashing. Use the LOCAL/TOTAL button the
advance the number and the ZERO button the move the number flashing to the left.
Enter the capacity of the scale and press the PRINT/ACCUMULATE button when
the scale capacity is displayed.
11. The display shows "LL-00". With no weight applied to the scale press the PRINT/
ACCUMULATE button.
12. The display shows "HH-01". Press the PRINT/ACCUMULATE button. The display
will show "00000" with the far right number flashing. Use the LOCAL/TOTAL button
to advance the number and the ZERO button to move the number flashing to the
left. Enter the value of the first load and press the PRINT/ACCUMULATE button
when the value of the first load is displayed.
13. The display shows "LL-01". Apply the first load to the scale. With the first load
applied to the scale press PRINT/ACCUMULATE button.
14. The display shows "HH-02". Press the PRINT/ACCUMULATE button. The display
will show "00000" with the far right number flashing. Use the LOCAL/TOTAL button
to advance the number and the ZERO button to move the number flashing to the
left. Enter the value of the second load and press the PRINT/ACCUMULATE
button when the value of the second load is displayed.
15. The display shows "LL-02". Apply the second load to the scale. With the second
load applied to the scale press PRINT/ACCUMULATE button.
16. The display shows "HH-03". Press the PRINT/ACCUMULATE button. The display
will show "00000" with the far right number flashing. Use the LOCAL/TOTAL button
to advance the number and the ZERO button to move the number flashing to the
left. Enter the value of the third load and press the PRINT/ACCUMULATE button
when the value of the third load is displayed.
17. The display shows "LL-03". Apply the third load to the scale. With the third load
applied to the scale press PRINT/ACCUMULATE button.