Preparation

Before determining the instrument errors, the instrument has to be
levelled using the electronic level.
The tribrach, the tripod and the underground should be stable and secure
from vibrations or other disturbances.
The instrument should be protected from direct sunlight to avoid thermal
warming.
It is also recommended to avoid strong heat shimmer and air turbulence.
The best conditions are early in the morning or with overcast sky.
Before starting to work, the instrument has to become acclimatised to the ambient
temperature. Approximately two minutes per °C of temperature difference from
storage to working environment, but at least 15 min, should be taken into account.
Even after adjustment of the ATR, the crosshairs may not be positioned exactly on the
centre of the prism after an ATR measurement has been completed. This outcome is
a normal effect. To speed up the ATR measurement, the telescope is normally not
positioned exactly on the centre of the prism. These small deviations/ATR offsets, are
calculated individually for each measurement and corrected electronically. This means
that the horizontal and vertical angles are corrected twice: first by the determined ATR
errors for Hz and V, and then by the individual small deviations of the current aiming.
Effects
Effects V Elimination
Hz
-
-
-
-
Automatically
with two face
corrected with
proper adjust-
measurement
ment
-
40