Note
60 mV range:
Thermovoltages occur in the event of temperature fluctua-
tion, which appear as additional voltage offset. It may be
necessary to repeat zero offsetting in order to achieve the
specified degree of accuracy.
Crest Factor Display
Crest factor is displayed for voltages in the V (AC+DC) function
along with the measured voltage value. The voltage value is mea-
sured simultaneously in a separate measuring circuit to this end,
and crest factor is displayed within a range of 1.0 to 11.0.
This value indicates the quality of the applied signal. The only pre-
requisite is a periodic signal with a valid frequency (see table).
Measuring range: 1.0
CF
Typical (not specified) maximum deviation
for U 5% of the measuring range:
CF 3.0
Frequency
02
10 to 70 Hz
02
70 to 440 Hz
05
440 Hz to 1 kHz
> 1 kHz:
Not valid
GMC-I Messtechnik GmbH
Measurements: V/Hz/dB – PQ/
11.0; resolution: 0.1
3.0 < CF 5.0
5.0 < CF 10.0
02
05
05
Not valid
Not valid
Not valid
Not valid
Not valid
– MHz/% – /nS/R
–
/
SL
– Temp –
/m – W/Wh – A/Hz/thd
27