Advanced Entry Calibration (continued)
Step
Action
Press the ESC Key.
Use the buttons to scroll to PAPER
SENSOR LEVEL (or if using
reflective media, REFL PAPER
N
LEVEL) and then press ENTER.
Use the buttons to set the Paper
value determined in Step M and
then press ENTER.
Press the ESC Key.
Scroll to GAP SENSOR LEVEL (or,
if using reflective media, MARK
SENSOR LEVEL) and then press
O
ENTER.
Set the Gap (or Mark) value
determined in Step M and then
press ENTER.
Press the ESC Key.
Scroll to EMPTY SENSOR LEVEL
P
and then press ENTER.
Set the Empty value determined in
Step M and then press ENTER.
Press the EXIT Key and, when
Q
prompted, press YES at the SAVE
CHANGES prompt.
Chapter 3 - Printer Operation
Displayed Message
PAPER SENSOR LEVEL
(0 - 255)
173
GAP SENSOR LEVEL
(0 - 255)
042
EMPTY SENSOR LEVEL
(0 - 255)
009
OFFLINE
Comment
This is the Paper value.
This is the Gap (or Mark) value.
This is the Empty value.
The procedure is finished, the
process almost complete.
30