Lifescan SureStepFlexx Operator's Manual page 79

Professional blood glucose management system
Table of Contents

Advertisement

5
Dry the lens area gently with a clean, soft, lint-
free cloth or tissue. Remove any lint.
6
Slide the closed test strip holder into the meter.
Push the holder until it clicks into place.
7
Turn on the meter. If an error appears, check
to make sure you installed the test strip holder
correctly.
8
If necessary, wipe the scanner lens with a soft
cloth dampened with water.
9
Perform a quality control test to verify that the
result is within the expected range.
Chapter 6: Care and Maintenance
67