• If the result is FAILED,
you may press
and choose from 1 to 3
comments. This will be
recorded with the failed
result. Press
the test strip, and repeat
the QC test.
Require test comment message entry for
failed QC test
You may be required to select a comment.
Refer to the SureStepPro Test Strip package
insert for information on troubleshooting a
failed QC test.
NOTE:
"Procedure Err" are not included in the statistical
calculations for any QC reports.
CAUTION:
the test. The system may not be functioning properly.
Do not use the system to test patient blood glucose
levels unless the control solution result falls within
the expected range.
If you have problems using the meter to test the
control solutions, or if the meter malfunctions,
call the LifeScan Healthcare Professional Line at
1 800 524-7226.
28
SureStepFlexx Meter Operator's Guide
Entr Note
, remove
Results tagged with the comment
If the control solution result fails, repeat
E n t e r N o t e s
FAILING QC
Pr oc ed u re Err
Re pe at ed Tes t
C lean ed M et er
M a t c h e d C o d e