MESSAGE
Shake vial and
apply (High,
Normal, Low)
control
Strip Lot
Expired
Upload
Overdue
Wrong Control
Type
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
90
SureStepFlexx Meter Operator's Guide
POSSIBLE CAUSE
Meter is ready for
control solution
application.
Test strip lot expired.
Data transfer not
performed within
specified time.
Type of control
(solution) entered does
not match test type.
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
SOLUTION
Gently shake vial. Touch a
drop of control solution to
center of pink test square.
• If white pad becomes sat-
urated, you have applied
too much solution. Repeat
application with new test
strip.
• If white patches or streaks
appear on confirmation
dot, you have not applied
enough solution. Repeat
application with new test
strip.
Enter and use new test strip
lot. Discard any remaining
expired test strips.
Perform data transfer session.
Check control type (low,
normal, high, linearity, or
unknown) and reenter
information.