If the white pad becomes completely saturated,
you have applied too much solution. Repeat the
application with a new test strip.
• Check the confirmation dot on the back of the
test strip to verify it has turned completely
blue, indicating an adequate amount of solution
was applied.
If white patches or streaks are visible, you have
not applied enough linearity solution. Repeat
the application with a new test strip.
• Wipe the linearity solution vial tip with a lint-
free cloth or tissue before replacing the cap.
CAUTION:
solution may cause out-of-range results.
9
Firmly insert the test strip all the way into the
test strip holder until it stops (the confirmation
dot should be facing down).
IMPORTANT:
strip, the test may start; however, you may receive an
out-of-range result.
54
SureStepFlexx Meter Operator's Guide
Applying too much or too little linearity
If you fail to completely insert the test