12. Hold Function

This function displays the hold weight data after averaging the weight data for a
specific period.
Useful to determine a live animal's weight.
Averaging time is selectable up to 9.9 seconds by 0.1 second step.
3 methods are available to start averaging; manual start, automatic start after
stable, and manual / automatic start.
Manual start is available with key switch or external input.
Serial interface commands are also available; averaging start, releasing the hold
data and outputting the hold state. Refer to "13.3.4. Commands for Hold Function".
Caution
This function can not be used under the setting CF00=1
Averaging can not start at a displayed value smaller than 0 +/- 5 digits.
Data when the display is over is not included for averaging.
When powered off, the hold is released automatically.
There is no peak hold function.
The display and the Data Output of Hold and Average
The weight display is blinking during the averaging period.
The output data in the averaging period is the actual weight at the time.
The weighing unit is blinking when the weight display is in the hold state.
The output data format of the hold weight data is the same as that of the normal
weight data except the header of stable state is "HD" in the response to the "RW"
or "RW,n" (n=1 or 2) command.
Relations to the Other Functions
If automatic accumulation (F20:21) and/or auto print (F37=2,3 F40=2,3) is set,
accumulation and/or data output is performed after determining the hold data.

12.1. Setting the Hold Functions

F27 determines the method of starting the average.
F27=1
F27=2
F27=3
*inhibit region
**stable detection
F28 determines the averaging time by 0.1second step. F28=0 holds the data at
averaging start.
12. Hold Function
Manual start: Starts the average and release with key switch operation.
Automatic start: After passing the inhibit region * and detect stable **,
starts the average automatically, releasing the data when the weight
returns to inhibit region.
Both Manual start and Automatic start.
0 +/- 5digits
Satisfied both F02 and F03