Dual Range
<First range> Specify the resolution, decimal point position and format.
<First range> Specify the weighing range.
<Second range> Specify the resolution.
<Second range> Specify the weighing capacity.
For the range function, refer to "5.3. Weighing Range Function".
5.2.2.

Specifying the Range and Unit

Step1
The range and unit of measure are displayed.
Range display :
Unit display
ENTER
ESC
5.2.3.

Specifying the Resolution, Decimal Point Position and Format

Step 2 The resolution will be displayed as d 0.1 with decimal point. The indicator
displays triangle 1 and the first unit selected at the previous step.
Specify the position of the decimal point with the
format (point or comma) with the +/- key. Specify the resolution with the
The decimal point format for serial data output is selected using the F-function
settings.
Press the ENTER key to store the parameters and proceed to the next step.
When pressing the ESC key, regardless of what is displayed, the indicator will
proceed to the next step without changing the parameters.
5.2. Calibration Procedure
SinGL : single range
dUAL : dual range
To change the range function, use the
The active unit is displayed. The unit can be changed;
such as kg or lb. Calibration is performed using the unit
displayed.
Press the
key to select a unit.
Press the +/- key to select a unit for calibration. The unit for
calibration (first unit) is displayed and the alternate unit (second unit)
is blinking.
The key to store the data displayed and proceed to the next step.
The key to proceed to the next step without changing the data.
Page 12
key
key. Specify the display
1
AD-4406 Weighing Indicator
key.