Calibration Error Code List - AND Weighing Indicator AD-4406 OP03 Instruction Manual

Table of Contents

Advertisement

5.6. Calibration Error Code List

Exiting from a Calibration Error
ESC key .... To return to the point where an error occurred. Retry the operation.
ESC key while pressing the ON/OFF key.
Error Code List
If an error has occurred during the calibration mode, the following code is
displayed.
Error code
With a multi-interval scale. The last division is set to maximum ( d-50 ).
Err 0
Therefore the next division can not be entered.
The resolution exceeds 40000. (Resolution = maximum capacity/ minimum
Err 1
division)
Reduce the maximum capacity or increase the minimum division.
The load cell output is too large or too small at zero calibration. Check the
Err 2
weighing unit and load cell. Refer to "3.5.Verifying Load Cell Output and Input
Err 3
Sensitivity".
Measuring calibration mass (or entering digital input), the value exceeded the
Err 4
maximum capacity. Reduce calibration mass (or digital input).
Err 5
The selected calibration mass is smaller than the minimum division.
The new input sensitivity is less than 0.2 µV/division. Increase the input
Err 6
sensitivity.
Refer to "3.5.Verifying Load Cell Output and Input Sensitivity".
Placing a mass on the weighing unit, the load cell output becomes a negative
Err 7
value. Check the load cell cable connections and the direction of load cell
mounting.
The load cell output exceeds the input range before the maximum capacity.
Adjust the zero balance referring to "3.5.Verifying Load Cell Output and Input
Err 8
Sensitivity". Replace with a load cell designed for a smaller output. Reduce the
maximum capacity.
The weighed value is out of the input range at zero calibration or span
Err 9
calibration. Check the weighing unit and cables.
Err 12
The previous weighing range is larger than next weighing range.
An incorrect mass is selected at the digital linearization function.
Err 13
Select a mass of the following relation. lnr 1 < lnr 2 < lnr 3.
G Err
An unacceptable value was selected in the gravity acceleration function.