Upper And Lower Limit-Line Testing - HP 8590 E-Series User Manual

Spectrum analyzers
Hide thumbs Also See for 8590 E-Series:
Table of Contents

Advertisement

Note
When entering a limit-line segment, the frequency, and amplitude values will
be listed as asterisks (***) until new values are entered. The new segment will
be listed last until both the frequency and amplitude values have been entered.
Once the frequency and at least one amplitude value are entered, the segment
will be sorted into the limit-line table according to frequency.
To edit an existing segment, use SELECT SEGMENT to specify the segment. Then use
SELECT FREQ , SELECT AMPLITUD , or SELECT TYPE to specify the column that you wish
to edit.
11. Press More 1 of 2 , then EDIT DONE when all values have been entered into the
limit-line table.
12. Press LMT TEST ON OFF so that ON is underlined. LIMIT PASS is displayed on the
spectrum analyzer screen if the measurement sweep is within the limit lines. LIMIT FAIL is
displayed if the measurement sweep is not within the limit lines.
P E A K
I O
W A V E
Figure 5-8. Upper and Lower Limit-Line Testing
Frequency
Amplitude Amplitude
Number
290 MHz
-60 dBm
2
3
295 MHz
-15 dBm
4
297 MHz
-15 dBm
5
300 MHz
6
303 MHz
305 MHz
-15 dBm
8
310 MHz
-60 dBm
9
400 MHz
-60 dBm
O
3 M H z
1
Lower
- 7 5 dBm
-75 dBm
Slope
-75 dBm
Slope
-29 dBm
Slope
-75 dBm
Slope
-75 dBm
Slope
-75 dBm
Flat
-75 dBm
Point
0 0 MHz
2 0 nlsec