Pulse Repetition
Interval (PRI)
100 ps
500 ps
1 ms
5 ms
10 ms
16.7 ms
33.3 ms
50 ms
100 ms
200 ms
249 ms
I

Using the Self-Calibration Routines with Option 105

The spectrum analyzer self-calibration routines (initiated by pressing CAL AMPTD or
CAL FREQ & AMPTD ) should be performed prior to using the Option 105 functions. Use the
following procedure to perform the self-calibration routines and to check the results of the
self-calibration routines.
Note
Be sure that the GATE TRIGGER INPUT connector (on the spectrum analyzer
rear panel) is not connected to anything while performing the spectrum
analyzer self-calibration routines.
1. Remove the cable from the GATE TRIGGER INPUT connector.
2. Connect the CAL OUT connector to the spectrum analyzer input connector with the
calibration cable.
3. Press a. Press either CAL FREQ & AMPTD (to perform the frequency and amplitude
self-calibration routines) or CAL AMPTD (to perform the amplitude self-calibration routine).
NEXT PAGE.
6. Verify that the number displayed for GATE, in the lower left corner, is between
0.98 and 1.0 (see Figure 4-30).
Pulse Repetition
Frequency (PRF)
10 kHz
2 kHz
1 kHz
200 Hz
100 Hz
60 Hz
30 Hz
20 Hz
10 Hz
5 Hz
4 Hz
Use the MAX HOLD trace function and take several measurement
Sweep Time (minimum)
21 ms
41 ms
201 ms
401 ms
2.01 s
4.01 s
6.7 s
13.4 s
20.1 s
40.1 s
80.2 s
100 s
Making Measurements 4-35