CAL FREQ & AMPTD routine before trying to store the correction factors. This message also
sets SRQ 110. (U)
CAL: FM SPAN SENS FAIL
The spectrum analyzer could not set up span sensitivity of the FM coil. (H)
CAL: GAIN FAIL
Indicates the signal amplitude is too low during the CAL AMPTD routine. This message also
sets SRQ 110. (H)
CAL: MAIN COIL SENSE FAIL
The spectrum analyzer could not set up span sensitivity of the main coil. If this
message appears, press
DEFAULT CAL DATA , and perform the CAL FREQ routine again. (H)
CAL: NBW 200 Hz notch amp failed
Indicates that the 200 Hz resolution bandwidth is not the correct shape for the calibration
routine. (H)
CAL: NBW 200 Hz notch failed
Indicates that the 200 Hz resolution bandwidth is not the correct shape for the calibration
routine. (H)
CAL: NBW 200 Hz width'failed
Indicates that the 200 Hz resolution bandwidth is not the correct bandwidth for the
calibration routine. (H)
CAL: NBW gain failed
Indicates that one of the resolution bandwidths is not the correct amplitude for the
calibration routine. (H)
CAL: NBW width failed
Indicates that one of the resolution bandwidths is not the correct width for the calibration
routine. (H)
CAL: PASSCODE NEEDED
Indicates that the function cannot be accessed without the pass code. For the DEFAULT CAL
DATA function, the pass code is setting the center frequency of the spectrum analyzer to
-37
Hz. (M)
CAL: RES BW AMPL FAIL
The relative insertion loss of the resolution bandwidth is incorrect. This message also sets
SRQ 110. (H)
CAL SIGNAL NOT FOUND
Indicates the calibration signal (CAL OUT) cannot be found. Check that the CAL OUT and
the spectrum analyzer input connectors are connected with an appropriate cable. If the
calibration signal is connected to the spectrum analyzer input but cannot be found, press
-37,
calibration signal still cannot be found, press (j-1, -37, IHz) and perform the CAL
FREQ or CAL FREQ & AMPTD self-calibration routines. This message also sets SRQ 110. (U)
and (H)
CAL: SPAN SENS FAIL
The self-calibration span sensitivity routine failed. This message also sets SRQ 110. (H)
CAL: USING DEFAUIX DATA
Indicates that the calibration data is corrupt and the default correction factors are being
the
used. Interruption of
9-8
If You Have A Problem
-37,
(Hz), (CAL], More 1 of 4 , More 2 of 4,
self-calibration routines or
an
error can cause this problem. (M)