Adjacent Channel Power Ratio (Acp) - HP 8590 E-Series User Manual

Spectrum analyzers
Hide thumbs Also See for 8590 E-Series:
Table of Contents

Advertisement

Adjacent Channel Power Ratio (ACP)

The leakage of a transmitter into adjacent channels can be measured quickly and easily.
Enter the channel spacing and channel bandwidth, and activate the adjacent channel power
measurement. The spectrum analyzer computes and displays the ACP ratio of both the lower
and upper adjacent channels, marking the higher of the two. The absolute carrier power is read
out and vertical lines on the display mark the channel bandwidth edges.
Selecting ACPGRAPH ON OFF (ON) computes a graph showing the adjacent channel power ratio
for the selected channel as a function of channel spacing, and disables the numerical adjacent
channel power display.
Normally, the spectrum analyzer measurement parameters are set automatically. Parameter
setting can be changed from automatic to manual control using PARAM AUTO MAN . The
following conditions must be maintained to make a valid rms measurement. If these conditions
are not met, errors of up to -2.5 dB can occur for noise-like signals.
Video bandwidth is at least 10 times the resolution bandwidth.
n
mode.)
Resolution bandwidth is less than or equal to 100 kHz.
n
Video averaging is OFF.
n
Neither MAX HOLD nor MIN HOLD trace mode is selected.
n
A wider dynamic range is available using the ADJ CHAN POWER extd function. This extended
range is measured by taking two different sweeps with different reference levels and
combining the trace data. The results are displayed in a 13 dB per division format. The
extended ACP function does not provide a continuous measurement mode.
Example: Measure the adjacent channel power of a signal.
1. Connect a signal to the spectrum analyzer INPUT 500.
2. Press
and enter the frequency of your input signal to place the signal at the
spectrum-analyzer center frequency.
3. Press [AMPLITUDEI] and adjust the reference level to bring the signal near the top of the
display.
will use the last entered values for channel bandwidth and spacing, if they are not
entered.)